Benchtop Terahertz Time-Domain Spectrometer TDS10XX

Poster THz spectrometer TDS1008 (pdf)
Support: Determination of material parameters from THz measurements (pdf)

TDS10XX

The benchtop THz spectrometer TDS10XX is easy to use and meant for laboratories where material characterization shall be performed in the THz region. The spectrometer comes with an internal sample compartment (which can be purged with nitrogen) or external fiber-coupled antennas. Both setups can be used to conduct transmission and reflection measurements with a collimated or focused THz beam. The TDS10XX is controlled via USB using a laptop with the pre-installed T3DS software package.

The following optional features are available:
  • SHR - Sample Holder for Reflection measurements
  • SHA - Sample Holder for Attenuated total reflection measurements
  • SHF - Sample Holder for Fiber coupled antennas
  • FSU - Fast Scan Unit with maximum scan rate of 100 Hz
  • IU - Imaging Unit for fiber coupled antennas
  • T2T - Theta-2-Theta angular scanning unit for fiber coupled antennas

Time domain THz spectrometer

Video: THz time-domain spectrometer TDS1008 with nitrogen purged sample compartment for transmission and reflection measurements. Additional fiber coupled THz antennas can be connected for external measurement setups.

TDS10XX data and price information

The performance and the price of the spectrometer depend on the femtosecond laser used and the chosen configuration. The best results can be achieved using a 800 nm laser source (TDS1008) and the internal sample compartment with a nitrogen purge.

General parameters
  • Maximum scan range*: 500 ps
  • Spectral resolution @ maximum scan range*: 2 GHz
  • Collimated THz beam diameter: 22 mm
  • Focused THz beam diameter @ 1 THz: ~ 1 mm
  • Electric power: 110 - 230 V, single phase
  • Dimensions (L/W/H): 90 cm / 60 cm / 25 cm
  • Weight: 90 kg

* Longer scan ranges / higher spectral resolutions are available upon request.

Downloads

TDS with 800 nm femtosecond laser


TDS1008 with sample compartment
  • Bandwidth: 0.05 - 4.5 THz
  • Dynamic range: ≥ 85 dB
  • Price: RFQ or send an e-mail
TDS1008-fo with fiber coupled antennas only
  • Bandwidth: 0.05 - 3.5 THz
  • Dynamic range: ≥ 70 dB
  • Price: RFQ or send an e-mail
TDS1008+f with sample compartment + fiber coupled antennas
  • Price: RFQ or send an e-mail

TDS with 1550 nm femtosecond laser


TDS1015 with sample compartment
  • Bandwidth: 0.05 - 1.5 THz
  • Dynamic range: ≥ 65 dB
  • Price: RFQ or send an e-mail
TDS1015-fo with fiber coupled antennas only
  • Bandwidth: 0.05 - 1.5 THz
  • Dynamic range: ≥ 65 dB
  • Price: RFQ or send an e-mail
TDS10015+f with sample compartment + fiber coupled antennas
  • Price: RFQ or send an e-mail

Optional parts for TDS10XX

The measurement capabilities of the THz spectrometers listed above can be extended by additional devices.

Prices apply to individual pieces, we grant a discount for larger quantities. Please ask!

> SHT - Sample Holder Transmission


The sample holder transmission consists of a plate, two sleds with focusing THz lenses, one adapter for large samples (at least 30 x 30 mm) with collimated THz beam and one adapter for small samples (at least 10 x 10 mm) with focused THz beam.
The SHT needs as support the sample holder base (SHB) which is not included, but comes with every TDS system.

Data sheet (pdf)
• Price: 1,250.- €

Photo SHT

SHT - Sample Holder Transmission

SHT on base structure

SHT on base structure

> SHR - Sample Holder Reflection


The sample holder reflection consists of a plate, two sleds with focusing THz lenses, one adapter for large samples (at least 30 x 30 mm) and one adapter for small samples (at least 10 x 10 mm). The angle of incidence or the SHR is 30° by default.
The sample holder for reflection measurements requires the base structure (SHB) which is not included, but comes with every TDS system.

Data sheet (pdf)
• Price: 1,450.- €

sample holder reflection

SHR -Sample Holder Reflection

sample holder reflection

SHR -front side

> SHA - Sample Holder Attenuated Total Reflection (ATR)


The sample holder for ATR measurements on strong absorbing liquids or solids with a refractive index lower then 2.5 consists of a mounted silicon prism with refractive index 3.416. The incidence angle of the THz beam on the prism base is 51.6 deg.
The sample holder for ATR measurements requires the base structure (SHB) which is not included, but comes with every TDS system.

Data sheet (pdf)
• Price: 2,450.- €

sample holder attenuated total reflection

SHA with lid

sample holder attenuated total reflection

SHA without lid

> SHB - Sample Holder Base


The sample holder base SHB is required for sample holder transmission (SHT), sample holder reflection (SHR) and sample holder attenuated total reflection (SHA). It comes with every TDS system and ensures the correct adjustment of the sample holders in the THz beam.

Data sheet (pdf)
• Price: 1,200.- €

SHB - sample holder base

SHB - Sample Holder Base

SHB with SHR

SHB with sample holder reflection (SHR)

> CUV - TPX CUVette for THz transmission measurements of liquids


The TPX Cuvette can be used for transmission measurements of liquids. For the best results we recommend using our Sample Holder Transmission (SHT) to position the TPX Cuvette properly in the THz beam path. The SHT comes with every TDS10XX Spectrometer but is not part of the cuvette.

• Material: TPX (Polymethylpentene)
• Complex refractive index n+i·k= 1.46 + i·8.5·10-4
• Dimensions: 50 mm x 20 mm x 41 mm (L x W x H)
• Thickness of TPX walls: 2 x 5 mm
• Liquid path length: 10 mm
• Time delay of empty cuvette: 15.3 ps
Data sheet (pdf)
• Price: 495.- €

TPX cuvette

TPX cuvette

Cuvette on base structure

Cuvette on base structure

> SHF - Sample Holder for Fiber coupled antennas


The sample holder for fiber coupled antennas consists of a base structure, one adapter for large samples (at least 30 x 30 mm) with collimated THz beam and one adapter for small samples (at least 10 x 10 mm) with focused THz beam. The SHF can be used for both, transmission and reflections measurements.
The SHF comes with every TDS system that was ordered with fiber coupled antennas.

Data sheet (pdf)
• Price: 1,500.- €

sample holder attenuated total reflection

SHF for a) collimated beam, b) focused beam in reflection, c) focused beam in transmission

> FSU - Fast Scan Unit


The Fast Scan Unit is used to change the laser beam path with a frequency of a few Hz in order to get a real time picture of the THz signal. It is typically used as an extension to simplify the signal optimization. Due to the limited scan range and time resolution it cannot be used to collect data at the same quality as with a slow scan measurement. However, the much faster data acquisition allows to change the optical or THz beam while watching the signal live.

•100 Hz @ 5 ps scan
•10 Hz @ 40 ps scan
Data sheet (pdf)
• Price: 9,200.- €

fast scan unit

FSU - Fast Scan Unit

fast scan unit

FSU with electronic box

> IU150 - Imaging Unit 150 mm x 150 mm


Imaging unit with x-y scan range 150 mm x 150 mm.
Larger imaging areas are available on request.
Can be used in reflection or transmission mode.
The software for THz imaging and visualization is included.
The imaging unit needs fiber coupled antennas outside the spectrometer.
Does not include a sample holder.

• Price: 8,500.- €

imaging unit in transmission

IU150 in transmission

imaging unit in reflection

IU150 in reflection

Imaging measurement in reflection

2 € coin, visible picture

2 € coin
visible image

2 € coin, THz amplitude image

THz amplitude / time delay
height profile image

2 € coin, integrated spectrum image

Integrated THz spectrum
specular reflectance image

> 2TPX - set of two mounted TPX lenses 1" diameter


Set of two mounted aspheric TPX lenses with 1" diameter for transmission measurements on small samples or for THz imaging.
Lenses with the shortest focal lnegth of 10 mm ( F* = 0.5) ensures a spatial resolution of 0.3 mm for frequencies > 1 THz.

• Available focal length: 10mm, 25mm, 32.5mm, 50 mm, 67 mm, ..
• TPX lens details you can find here..
Spatial resolution measurement (pdf)
• Price: 420.- € / set

TPX lens

TPX-D25.4-f10

 Photo TPX lens

Set of two mounted TPX lenses

> T2T-m - Theta-2-Theta angular scanning unit, manually operated


The incidence angle theta of the Theta-2-Theta reflection unit is manually set. The reflection angle theta is automatically set equal to the incidence angle theta (specular reflection).
Needs fiber coupled antennas outside the spectrometer.

Data sheet (pdf)
• Price: 3,500.- €

T2t-m unit

T2T-m

> T2T-a - Theta-2-Theta angular scanning unit, automatically operated


The reflection angle theta is automatically set equal to the incidence angle theta (specular reflection).
An automatic angular scan is possible using the spectrometer software T3DS.
Needs fiber coupled antennas outside the spectrometer.

Data sheet (pdf)
• Price: 7,000.- €

T2t-a unit

T2T-a

Measurement examples using the T2T-a: Angular scan of a reflected, collimated THz beam

Object: Al grating

Object: ruled Al grating
1 mm groove width

Angular scan for reflected beam on ruled grating

Angular scan for reflected beam
on ruled grating